Siemens Mobilett ll Mobile Portable X-Ray Xray Imaging

Available

Siemens Mobilett ll Mobile Portable X-Ray Xray Imaging


$2,999.00

Loyalty program By buying this product you can collect up to 299 loyalty points. Your cart will total 299 points that can be converted into a voucher of $239.20.


30 other products in the same category:

Good Condition, with Instruction Booklet, Exposure table, standard Values.

Tube Housing: Model# 1158815 S/N

Generator: Model# 902342 S/N 3047 S11

Tube Manufac: Siemens Sept 1997 Siemens ID Tag 1631           
Specifications:
Year of Manufacture: NOVEMBER 2002
Condition:
The equipment is deinstalled and in storage.

Last inspection was on 13th December 2010, in very good working Condition, no error.

Please here is the summary of the inspection result:

Session Summary

Performance Report
Half Value Layer Minimum HVL @ 80 kVp is 2.9mm Al 3.2 Pass
kVp Accuracy Maximum deviation should be within 10% of the selected value. 2.6% Pass
Accu-kV Waveform User defined Pass
Reproducibility Coef Var. must not be greater than 0.05 and each of the 10 exposures must be within 15% of the mean. 0.042 Pass
Linearity mR/mAs at any 2 consecutive tube current setting must not differ by more than 0.10 of there sum. 0.018 Pass

No customer comments for the moment.

Only registered users can post a new comment.